Flame Photometers
|
|
Model FPH is a micro-controller based
instrument designed to provide automation in operation,
measurements and end-result presentation. It does the estimation
of Sodium (Na), Potassium (K), Lithium (Li), and Calcium (Ca) in
single aspiration of a sample. For user’s convenience, the
instrument offers two measuring modes: W Low Concentration and
(ii) High Concentration. There is facility of curve fitting
correction for non-linear emission characteristics of element,
at higher concentration, store (save) Setup for frequently used,
measurement (in a battery-backup memory) and the same can be
recalled whenever required with a press of a button. Also
re-standardization using single standard to minimize the effect
of any unforeseen drift without going for the full
recalibration. 4 Line 20-character LCD readout provides easy
user interface & presentation of results. A Centronics printer
port provides print out of results on EPSON compatible Dot
Matrix Printer; supply of Printer is an optional. |
|
|
| Salient Features |
-
Micro-controller based automation for ease
of operation.
-
Determination up to four elements with single
aspiration (Na, K, Li, and Ca).
-
Curve Fitting Correction for non-linear
emission characteristics of element, at higher concentration;
maximum five point calibration
-
Storing facility for Set-ups and results (date
wise & batch wise, maximum 700 results); the same can be
recalled whenever need arise.
-
Four Line 20-character LCD readout
-
Facility for re-standardization with single
standard.
-
Multi standard graph in dot matrix printout
-
Built-in real time clock for date and time of
analysis.
-
Centronics printer port provided for print out
of results (simultaneous with analysis or recalled) on EPSON
compatible Dot Matrix Printer; supply of Printer is optional
-
Printout facility for individual sample, batch
sample, total day's samples, or entire results in the memory
(700 max.)
-
Air compressor with built-in air regulator and
air filter (SYSTRONICS 126)
-
PC Link software for transferring data to
computer & manipulation of data in computer using RS-232
(Optional)
|
|
|
|
Technical Specifications
|
|
Elements
|
Low Conc. Mode (Max)
Without Dilution
|
High Conc. Mode (Max)
Without Dilution
|
|
Na |
10 ppm. |
100 ppm. |
|
K |
10 ppm. |
100 ppm. |
|
Li |
2 ppm. |
50 ppm. |
|
Ca |
100 ppm. |
300 ppm. |
-
Suitable
dilutions should be made for solutions of higher
concentrations than given in the above table.
-
A curve fitting
software is provided for the High concentration Mode.
Accuracy: ± 2.0% of Full scale
-
For reliable
results the concentrations of the diluted solutions should
not be lower than the 1 /10th of the selected Range.
|
|
|
|
|
Full Scale Sensitivity |
|
Na : 2 ppm |
K : 1 ppm |
Li : 1 ppm |
Ca : 30 ppm |
|
|
|
|
Minimum Detective Limit |
|
Na : 0.2 ppm |
K : 0.1 ppm |
Li : 0.1 ppm |
Ca : 3 ppm |
|
-
FILTERS (10 nm Typical) : Na and K supplied;
Li and Ca (optional)
-
REPRODUCIBILITY : Low Conc.: ± 1 % f.s.; High
Conc. : ± 2% f.s.
-
MINIMUM SAMPLE : Approx 3 ml per element (at
Avg. Time of 4 sec.)
-
AVERAGING : 2 to 15 seconds, selectable
-
ASPIRATION TIME : (5 Sec + Avg. Time) per
element + 4 Sec.
-
OPERATING AIR PRESSURE : 0.45 kg / cm2
(typical)
-
AIR COMPRESSOR : With
built - in air regulator and air filter to deliver stable and
moisture / oil free air supply.
-
FUEL GAS : LPG
-
POWER SUPPLY : 230 Vac 10%, 50 Hz.
|
|
Standard
Accessories: The system consist of Main Unit with one each
of Na & K narrow band filters, a compressor with built - in air
regulator and air filter.
Optional: Li and Ca filters and printer |
|
 |