Model
TD - 01 is an accurate and elegant instrument designed for
conducting nephelometric and turbidity measurements of suspended
particles and colloids in solution. The principle of operation
of this instrument is based on the well-known Tyndall effect. A
beam of light passing through a turbid liquid being tested,
scatters the light which, in turn, is collected at right angles
by a Photocell and indicated in a display. The amount of
scattered light is proportional to the turbidity of the solution
under test. The salient features of the instrument are its
simplicity of operation, accuracy and reliability. The
instrument is calibrated directly in N.T.U.'S. The power supply
to the 6W lamp is stabilized and the amplifier uses IC-
circuitry. The components, electronic as well as optical, are
mounted in such a way that they are easily accessible for
service. The instrument operates on 230V, 50 Hz nominal AC
supply and is housed in an elegant two -tone cabinet. The
applications of turbidity measurements are vast -in Municipal,
Public Health, Sewage Works and in Industrial Establishments for
determining quality of water/waste water. |
|
Specifications
|
|
Ranges |
Four ranges (a) 0 - 1 NTU (b) 0 - 10 NTU
(c) 0 - 100 NTU (d) 0 - 1000 NTU |
|
Accuracy |
Within ±3% of FSD in 0-1 and 0-1000 NTU ±2%
of FSD in 0-10 and 0-100 NTU ranges |
|
Reproducibility |
Within ±3% of FSD |
|
Amplifier |
IC - circuitry |
|
Detector |
Visible range Photodetector |
|
Calibration |
With Formazine standard solution |
|
Accessories |
Flat bottom test
tubes - (25mm dia.) 4 Nos., Cell riser, Light Shield,
Instruction Manual |
|
Power Supply |
230 Volts ± 10%, 50Hz, AC |
|