Digital Photoelectric Colorimeters
|
|
‘LOBAlife’ Photoelectric Colorimeters
Models COL - 20 & COL - 15 are versatile models which can
effectively serve both the Chemical and the Clinical groups in
fields such as Biochemistry, Agriculture, Veterinary, Dyestuffs
& Pharmaceuticals They can measure Percentage Transmittance (%T)
and Absorbance (ABS). Improved optics enable measurement with
only 1ml solution. A wide range of wavelength, 400 to 700 nm, is
covered with the help of 8 optical filters mounted on a turret.
Additional features of COL - 15 are measurement of concentration
(Cons), K-Factor & accuracy to 3 decimals. The unit accepts 10ml
square cuvetts for precise measurement. Measurement can be made
with 300 ul in 1ml cuvett or 1ml in 4 ml test tube / cuvette.
Facility is provided to use narrow band interference filters
(optional) of user's choice for specific application to achieve
precise results. |
|
|
|
COL - 20 |
-
%
T & ABS (O. D.)
-
Sample volume needed is only 1ml in 4 ml
test
tube
-
3 digit led display.
-
Eight filters on a turret to cover 400-700 nm
-
Hermetically sealed photodiode adds ruggedness
|
|
COL - 15 |
-
%
T, ABS, CONC and K-factor
-
Accepts square and round test-tube
-
Sample volume needed is only 300 ul in 1 ml
cuvette 1ml in 4 ml cuvette test tube
-
Eight filters on a turret to cover 400-700 nm
-
Facility to drop in a narrow band interface
filter for specific applications
-
Hermetically sealed photodiode adds ruggedness
-
Improved optics to minimise operational
aberrations
-
Ergonomically laid out controls
-
Soothing 4-digit LCD display
|
|
|
|
SPECIFICATIONS |
COL -15 |
COL - 20 |
|
Wavelength Range |
400-700nm with 8 optical filters. |
400nm to 700nm with 8 optical filters. |
|
Filter's peak wavelength |
420nm, 440nm, 490nm, 520nm, |
420nm, 440nm, 490nm, 520nm, |
|
(typical) |
540nm, 570nm, 600nm, 700nm. |
540nm, 570nm, 600nm, 700nm. |
|
Measuring Modes |
%T, ABS, CONC and K-factor. |
%T & ABS |
|
Photometric Resolution |
0.1%T, 0.001 ABS upto 1.999 ABS (O.D.) |
1% T & 0.01 in ABS upto 1.99 ABS (O.D.) |
|
Display |
4-digit LCD. |
3-digit LED. |
|
K-Factor Range Display |
0 to 1999. |
- |
|
Concentration |
0 to 2 times the absorbance |
- |
|
setting limit |
readout (with decimal point removed). |
- |
|
Sample Volume (min.) |
1ml in 4ml cuvette / Test tube. |
1ml in 4ml Test tube. |
|
Sample Holder |
(i) 10 mm square cuvette.
(ii) 12 mm round flat bottom test-tube. |
12 mm round flat bottom test-tube. |
|
Source |
6V, 3W pre-focused |
lamp. 6V, 0.3A |
|
Detector |
Hermetically Sealed photodiode. |
Hermetically Sealed photodiode. |
|
Power |
230V ± 10%, 50 Hz. |
230V ± 10%, 50 Hz. |
|
Weight |
2.5 Kg. (approx.) |
2.5 Kg. (approx.) |
|
|
|
Optional |
|
Interference filter |
500, 546, 578, 630nm etc. typical BW of 10nm. |
|
Sample Volume (min.) |
300 ul in 1ml cuvette |
|
|
|
|
COL -15 / COL - 20 use 8 fitters with these characteristics. |
|
Filter No |
Peak (nm) |
Passband (nm) |
Colour |
|
1. |
420 |
380-510 |
Violet |
|
2. |
440 |
400-530 |
Blue |
|
3. |
490 |
460-540 |
Blue/Green |
|
4. |
520 |
490-560 |
Green |
|
5. |
540 |
510-590 |
Yellow/Green |
|
6. |
570 |
570-590 |
Yellow |
|
7. |
600 |
570-700 |
Orange |
|
8. |
720 and above |
630 and above |
Red and above |
|
|
Zero No for COL - 15 : OPEN, No filter. This
position is set when an optional Interference Filter is to be
used. |
|
 |
|
|
|
|
|
|
|