‘LOBAlife’ SPECTROPHOTOMETERS VSP06 & VSP66 is a
set of micro controller based new standards in accuracy,
reliability and time saving. Wide spectral coverage and high
resolution is achieved by using a powerful light source and a
grating based monochromator. A conveniently placed control
enables the user to quickly orient the grating to provide
monochromatic light of desired wavelength. The electronic
processing circuits and power supplies are solid-state/LSI-based
to ensure high accuracy, stability and long life. Two different
models, catering to the varying needs of end-users, are
presented here. |
|
|
| Model VSP06 |
-
4-Digit digital display for Transmission, Absorbance,
Concentration and K-factor.
-
Wavelength range
340-960nm
-
Wavelength
readout - 5 nm step (Dial type)
-
Wavelength accuracy ± 5nm
-
Band width - 5nm
-
Sample Holder - Two position holder for 10 mm
cuvette.
-
Optics-600 lines/mm grating based
monochromator.
|
|
|
| Model VSP66 |
-
4 Digit digital display for Transmission, Absorbance,
Concentration and K-factor.
-
Wavelength range
340-990nm
-
Wavelength
readout - 3 Digit display with 1 nm step
-
Wavelength accuracy + 2nm
-
Optics-600 lines/mm grating based
monochromator.
-
Band width - 5nm
-
Sample Holder - Two
position holder for 10 mm cuvette.
|
|
|
|
|
|
‘LOBAlife’ VSP67 is a micro controller-based
Single Beam Scanning ,Visible Spectrophotometer. It provides
operating modes likes Single Wavelength, Multi Wavelength,
Spectrum scan & Concentration and measuring modes like %
Transmittance, Absorbance and Concentration by standard &
factor. The unit offers wavelength range from 340nm to 999nm,
wavelength setting resolution of lnm, wavelength setting
accuracy of ± 2 nm, and the spectral bandwidth of 5 nm. The
grating spectral order interference is eliminated by a set of
automatically positioned order cut off filters.
The unit is equipped with user friendly
software. A 20 characters & 2 line alphanumeric LCD Display
along with 16 keys enable the user to select, set and operate
the unit with ease. All the measurement results are available on
the display except for the spectrum scan. A commonly used EPSON
compatible dot matrix printer prints the scans besides providing
hard copy of the measurements. Added application software
provides data processing like Peak Pick (Absorbance peak) of the
spectrum scan. Concentration measurement with curve fit (up to 5
standards) is provided. The unit can be connected to a PC
through an optional RS-232 interface to provide display of
spectra on monitor. This further adds features like: Spectrum
Storage/Retrieve, Peak-Pick/Point-Pick etc. The sample
compartment provides a TWO POSITION cuvette holder of I Omm path
length, as a standard facility. Minimum sample volume is 1 ml in
4ml cuvette. Single position 40mm/50mm path length cell holder,
factory installed, is optional. |
|
|
| Model VSP67 |
-
2 Line 20 Char. LCD Digital display for Transmission,
Absorbance, Concentration, K-factor, Multi Wavelength & Scan
(Low, Medium & Fast).
-
Wavelength range
340-999nm
-
Wavelength
readout 3 digit display with 1 nm step
-
Wavelength accuracy ± 2nm
-
Optics-6001ines/mm grating based
monochromator. Band width - 5nm
-
Sample Holder - Two position holder for 10 mm
cuvette.
-
Optional - Four
Position, Auto Sample Holder Port.
|
Click
here to view the comparison chart |
|
|
|
‘LOBAlife’ UV - VIS SPECTROPHOTOMETER UV17 is a
micro controller based single beam scanning instrument. It
provides three measuring modes ; Percent Transmission (%T),
Absorbance (ABS) and Concentration (CONC). The available
operating modes are Single-Wavelength, Multi-Wavelength,
Spectrum Scan and Time Scan. It covers 200-1100nm and offers 2nm
bandwidth. Combination of user friendly software, 20 x 4
alphanumeric LCD display and 18 soft-touch keys provide a stand
alone operation. All measurement results are available on the
display except for the Spectrum Scan. A printer prints the scan
besides providing hard copy of measurements. RS-232 is optional
to enhance data processing ability through PC. |
|
|
| Model UV17 |
-
LCD display 4 Line 20 characters for Transmittance,
Absorbance, Concentration by K-factor & Multi-standards.
-
Operating
modes - Single Wavelength, Multi Wavelength, Spectrum Scan and
Time Scan.
-
Optics - Single beam 1200 lines/mm grating based
-
Wavelength range - 200-1100nm
-
Wavelength readout - at 0.1 nm interval on LCD
display.
-
Wavelength accuracy - ±1 nm
-
Bandwidth - 2nm
-
Light source -
Tungsten halogen lamp, Deuterium Lamp.
-
Detector - wide range
silicon photo diode.
-
Sample Holder - Auto.
Five position holderfor I Omm cuvette.
|
|
|
|
|
|
PC OPERATED UV-VIS SPECTROPHOTOMETER UV19
has the versatility to meet the wide measurement needs of both
industries and research institutions at economical level.
The unit offers multi measurement modes : % Transmission,
Absorbance and Concentration. Special software, supports
Multi-standard Calibration with linear or nonlinear curve
fitting for concentration measurements, besides the K-factor
method.
Operation modes available are : Single Wavelength,
Multi-Wavelength, Wavelength Scan and Time-Scan. The PC
(Personal Computer) and application software enhance the data
processing ability and offer : Peak-Pick, Point-Pick, Expansion
of Spectra, Overlaying of Spectra, 1 ‘to 4th Derivatives,
Averaging of Scan, Subtraction of two Scan’s, Concentration
Calculations, Spectrum Store / Recall etc. (Pentium 700 MHz or
above is suitable)
HIGHLY STABLE BEAM
SYSTRONICS field proven optics combined with in-house designed
sophisticated advanced electronics circuits, and the digital
precision of the PC provides a highly stable beam.
AUTOMATIC BASE LINE CORRECTION
The unit automatically scans the blank/reference and stores its
spectrum in memory to give the base line correction when a
sample is scanned.
USER FRIENDLY AND MENU DRIVEN OPERATION
The instrument is operated through user friendly software. This
enables the user to select, set and operate the unit, in any of
the modes with ease. A CRT monitor displays all the menus, and
guides you through each step of operation and then displays the
results.
EDITING/SAVING THE CALIBRATION CURVE FOR
CONCENTRATION MEASUREMENT
Facility is provided for post - deletion of standards from a
selected calibration curve for concentration measurement to get
another more suitable curve for a given application. Standard
curve can be saved forfuture use.
SELF DIAGNOSTICS AND AUTO WAVELENGTH
CALIBRATION
Each time the unit is turned ‘on’ it checks its electronic sub
systems to ensure correct performance. Further, the unit goes
through a wavelength calibration routine to ensure wavelength
accuracy. |
|
|
| Model UV19 |
-
PC Based UV- Visible Spectrophotometer for Transmittance,
Absorbance, Concentration by Kfactor & Multi-standards.
-
Operating
modes - Single Wavelength, Multi Wavelength, Spectrum Scan and
Time Scan.
-
Optics - Single beam 1200 lines/mm grating based
-
Wavelength range - 200-999.9nm
-
Wavelength readout - at 0.2nm interval on
Monitor display.
-
Wavelength
accuracy-+1 nm
-
Band width -2nm
-
Light source -
Tungsten halogen lamp, Deuterium Lamp.
-
Detector- wide range
silicon photo diode.
-
Sample Holder- Manual
Four position holder for 10mm cuvette.
|
|
|
|
|
|
DOUBLE BEAM UV-VIS SPECTROPHOTOMETER UV21 (PC BASED) WITH
VARIABLE BANDWIDTH |
|
Versatile instrument for Research and
Industrial Laboratories. The instrument is designed with
precision and technique for accurate qualitative and
quantitative, Spectrophotometric analysis in Agricultural,
Chemical, Pharmaceutical, Medical, Bio-technological, Forensic
etc. fields. The PC based operation is user friendly for
beginners and gives enough scope for data handling to
experienced scientists.
The double beam optics with holographic
grating, in-house designed, sophisticated and advanced
electronic circuit and the digital precision of the PC enable
the instrument to perform at optimum level. A ‘high sensitivity
Photo-Multiplier Tube (PMT) used as a ,,detector offers high
wavelength resolution for identification of sharp peaks. The
bandwidth may be Selected from 0.5 nm to 6.0 nm (at an interval
of 0.1 nm) for the entire wavelength region. Variable ‘bandwidth
facility gives advantage to select lower ,bandwidth for sharp
peaks and higher bandwidth wherever possible to reduce the
signal to noise ratio. The computer controlled electronic
circuit moves a stepper motor driven screw-rod mechanism for
‘accurate wavelength setting with resolution of 0.1 nm Automatic
positioning five position 10 mm path length cuvette holder for
samples gives ease of operation. Also option is available for
single position, cuvette holder up to 100mm path length cuvette. |
|
|
|
Measurement modes available are : |
Operating modes available are : |
|
|
-
Single Wavelength
-
Multi wavelength
(Max. 15 wavelengths)
-
Scan (with multi Scan
facility) Max. 5 Scans
-
Time Scan (with
derivative facility) / Kinetic*
|
|
|
|
The PC and the specially developed application software
enhance the data processing ability and offer : |
-
Peak Pick/Point Pick/ Valley Peak
-
Expansion /Compression of Spectra
-
1st, 2nd, 3rd and 4th derivative
-
Averaging of Scans
-
Subtraction of two scans
-
Three levels of smoothening of spectraRatio,
Corrected ratio, two point and three point Net Absorbance,
useful for analysis of DNA/Protein.
* Recent Pentium PC’s with-SVCA card.
|
|
|
|
USER FRIENDLY WINDOW BASED AND MENU DRIVEN
OPERATION
The video monitor displays all the menus and guides through each
step of operation and then display the results. The graphical
and tabular data displayed on the monitor, can be stored on hard
disk or floppy disk and hard copy can be taken on dot-matrix,
ink-jet or Laser printer with graph size selection.
AUTOMATIC BASE LINE CORRECTION
Advance digital signal processing (DSP) technique is used for
automatic base line correction
SELF-DIAGNOSTIC AND AUTO WAVELENGTH CALIBRATION
Each time the unit is turned on the system checks the electronic
sub-systems to ensure correct performance, band-width of optical
system & sample holder positioning. |
|
|
| Model UV21 |
-
PC Windows Based
-
Modified
Czerny-Turner Monochromator geometry for better aberration
correction. Holographic diffraction grating with 1200
lines/mm.
-
Measuring Modes - Transmission, Absorbance and Concentration
by K-factor or by Multi-standards, Scanning & Time Scanning.
-
Optics : Double beam
-
Wavelength range: 190-1 000nm
-
Wavelength readout :
0.1 nm interval
-
Wavelength accuracy :
±0.5nm
-
Bandwidth: 0.5nm to
6nm (Variable)
-
Photometric range : ±
4 Abs.
-
Light source :
Tungsten Halogen Lamp/Deuterium Lamp.
-
Detector: Photo
Multiplier Tube. (PMT)
-
Sample Holder:
Automatic five position for sample and one fixed position for
reference for 10 mm cuvette
|
|
|
| Model UV22 |
-
PC Windows Based
-
Optics : Double beam
-
Wavelength range: 200-1100nm
-
Wavelength readout : 0.1 nm interval
-
Wavelength accuracy: ±0.5nm
-
Band width : 2nm
-
Photometric range: ±
4 Abs.
-
Light source :
Tungsten halogen lamp / Deuterium Lamp.
-
Detector : Dual
Si-Photo Diode
-
Sample Holder :
Automatic five position for sample and one fixed position for
reference for 10 mm cuvette.
-
Modified
Czerny-Turner Monochromator geometry for better aberration
correction. Holographic diffraction grating with 1200
lines/mm.
-
Measuring Modes -
Transmission, Absorbance and Concentration by K-factor or by
Multi-standards, Scanning & Time Scanning.
|
|
|
|
|
|
MODEL
UV23 |
|
Stand alone, Graphic LCD Display, Double beam optics : Modified
Czerny - Turner Monochromator geometry for better aberration
correction. Holographic diffraction grating with 1200 lines/mm
blazed at 250nm. |
|
|
|
Photometric |
|
Range |
±2.5 Abs |
|
Accuracy |
±0.005Abs at 1.0 Abs |
|
Repeatability |
±0.002Abs at 1.0 Abs |
|
Wavelength |
|
Range |
200 to 1100.00 nm |
|
Accuracy |
±0.5nm |
|
Repeatability |
±0.2nm |
|
Resolution |
0.1 nm |
|
Bandwidth |
<2.0 nm |
|
Stray Light |
Less than 0.05%T at 220nm and 340nm |
|
Baseline Correction |
Automatic Baseline correction |
|
Baseline Flatness |
Within ±0.003 ABS (excluding noises) |
|
Detector |
Photo Diode |
|
Click here to view the comparison chart |
|
|
|
|